Hapter, Roubleshooting – Labconco Coated Steel, Fiberglass and PVC Blowers 7183200 User Manual
Page 47
43
Product Service 1-800-522-7658, International 816-333-8811
C
HAPTER
8
T
ROUBLESHOOTING
Refer to the following table if your blower fails to operate
properly. If the suggested corrective actions do not solve your
problem, contact Labconco for additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Remote blower won’t
operate.
Wires not connected
at junction boxes or
switches.
Check connection of switches.
Check connection to control box
on top of unit.
Circuit breakers
tripped in building
electrical supply.
Reset circuit breakers.
Blower wiring is
disconnected.
Belt broken.
Blower motor is
defective.
Inspect blower wiring and switch.
Replace belt.
Replace blower motor.
Contamination outside
of fume hood.
Fume hood has
improper face
velocity
Have fume hood re-certified and
check remote blower exhaust
system. Hood should have
average face velocity of 60-100
fpm depending on application.
Remote blower has
excessive vibration.
Improper motor
mount.
Review Chapter 2: Prerequisites
and Chapter 3: Getting Started.
The blower should be mounted on
vibration isolators or vibration
mounting pads to isolate vibration.
- Coated Steel, Fiberglass and PVC Blowers 7183100 Coated Steel, Fiberglass and PVC Blowers 7183000 Coated Steel, Fiberglass and PVC Blowers 7182300 Coated Steel, Fiberglass and PVC Blowers 7182200 Coated Steel, Fiberglass and PVC Blowers 7182100 Coated Steel, Fiberglass and PVC Blowers 7182000 Coated Steel, Fiberglass and PVC Blowers 7181700 Coated Steel, Fiberglass and PVC Blowers 7181600 Coated Steel, Fiberglass and PVC Blowers 7181500 Coated Steel, Fiberglass and PVC Blowers 7181400 Coated Steel, Fiberglass and PVC Blowers 7181300 Coated Steel, Fiberglass and PVC Blowers 7181200 Coated Steel, Fiberglass and PVC Blowers 7181100 Coated Steel, Fiberglass and PVC Blowers 7181000 Coated Steel, Fiberglass and PVC Blowers 7180900 Coated Steel, Fiberglass and PVC Blowers 7180800 Coated Steel, Fiberglass and PVC Blowers 7180700 Coated Steel, Fiberglass and PVC Blowers 7180600 Coated Steel, Fiberglass and PVC Blowers 7180500 Coated Steel, Fiberglass and PVC Blowers 7180400 Coated Steel, Fiberglass and PVC Blowers 7180300 Coated Steel, Fiberglass and PVC Blowers 7180200 Coated Steel, Fiberglass and PVC Blowers 7180100 Coated Steel, Fiberglass and PVC Blowers 7180000 Coated Steel, Fiberglass and PVC Blowers 7071900 Coated Steel, Fiberglass and PVC Blowers 7069700 Coated Steel, Fiberglass and PVC Blowers 7069600 Coated Steel, Fiberglass and PVC Blowers 7069500 Coated Steel, Fiberglass and PVC Blowers 7069400 Coated Steel, Fiberglass and PVC Blowers 7069300 Coated Steel, Fiberglass and PVC Blowers 7069200 Coated Steel, Fiberglass and PVC Blowers 7069100 Coated Steel, Fiberglass and PVC Blowers 7069000 Coated Steel, Fiberglass and PVC Blowers 7068900 Coated Steel, Fiberglass and PVC Blowers 7068800 Coated Steel, Fiberglass and PVC Blowers 7068700 Coated Steel, Fiberglass and PVC Blowers 7068600 Coated Steel, Fiberglass and PVC Blowers 7068500 Coated Steel, Fiberglass and PVC Blowers 7068400 Coated Steel, Fiberglass and PVC Blowers 7068300 Coated Steel, Fiberglass and PVC Blowers 7068200 Coated Steel, Fiberglass and PVC Blowers 7068100 Coated Steel, Fiberglass and PVC Blowers 7068000