beautypg.com

Exhaust transition adaptor, Accessory for perchloric acid applications, Wash rings – Labconco Coated Steel, Fiberglass and PVC Blowers 7183200 User Manual

Page 45: Hepa filter packs, Charcoal filter packs

background image

41

Product Service 1-800-522-7658, International 816-333-8811

Chapter 7: Modifying Your Blower


Accessory for Basic 47 Hoods

Exhaust Transition Adaptor

The exhaust transition adapts to 7" and 10" rectangular outlet on
Basic 47 Hoods, model series 22473 and 22475, to receive 10"
diameter PVC duct.

Nominal Diameter/Inches

10

Catalog Number

22648

Shipping Wt./lbs.

5


Accessory for Perchloric Acid Applications

Wash Rings

Wash rings are suited for use in Perchloric acid duct systems.
Each features a wide-angle conical spray nozzle and wash water
connector nipple fabricated into a PVC coupling for use with 10"
or 12" nominal duct.

Nominal Diameter/Inches

10

12

Catalog Number

47460

47461

Shipping Wt./lbs.

5

6


Accessories for Pathogens, Organic Vapors and Odor Control
Applications

HEPA Filter Packs

High Efficiency Particulate Air Filter for non-radioactive
particulate and pathogenic applications. Rated for 1000 CFM
airflow with initial 1.0" static pressure drop. Replaceable HEPA
filter media removes 99.97% of all particles 0.3 micron or greater.
Furnished with clamping frame and duct connections. Unit
measures 28" x 28" x 25" high.

Charcoal Filter Packs

Activated Charcoal Filter for non-radioactive organic vapors and
odor control are rated for 1000 CFM airflow with 0.2" static
pressure drop. Unit measures 28" x 28" x 25" high.

Nominal Diameter/Inches

8

10

12

HEPA Filter

Catalog Number

22400

22401

22442

Shipping Wt./lbs.

100

100

100

Charcoal Filter

Catalog Number

22430

22431

22441

Shipping Wt./lbs.

100

100

100



This manual is related to the following products: