beautypg.com

Mounting support requirements, Electrical power requirements, Space requirements – Labconco Coated Steel, Fiberglass and PVC Blowers 7183200 User Manual

Page 10

background image

6

Chapter 2: Prerequisites

Product Service 1-800-522-7658, International 816-333-8811

Mounting Support Requirements


You must provide vibration isolators, vibration mounting pads,
and/or a roof curb support for proper mounting of the blower.
Vibration isolators or vibration mounting pads are available from
many sources such as a local industrial supply company.
Labconco recommends supporting the blower with 5/16" diameter
mounting hardware.

Electrical Power Requirements


The blower wiring should be terminated at the motor end plate and
wired according to the specific voltage and terminations on the
motor. Locate the specific wiring voltage for your blower motor in
the chart in Chapter 4: Performance Data and Safety Precautions.

Space Requirements


The dimensions for the different models are shown in Appendix B:
Blower Dimensions
.



This manual is related to the following products: