Location requirements – Labconco PuriCare Procedure Stations 31261 Series User Manual
Page 9
Chapter 2: Prerequisites
Product Service 1-800-522-7658
3
Original instructions
Location Requirements
Note: The procedure station should be located away from traffic patterns,
doors, fans, ventilation registers, fume hoods and any other air-handling
devices that could disrupt its airflow patterns. All windows in the room
should remain closed. Figure 2-1 shows the preferred location for the
procedure station.
Preferred location
Air register blocked or redirected
to prevent cabinet disruption
Alternate location
Main Disconnect Device
(Power cord plug in back
of the electronics module).
Do not position the unit so that it is difficult to operate the main
disconnect device.
Ne placez pas l'appareil de sorte qu'il est difficile de faire fonctionner le
dispositif principal de déconnexion.
Figure 2-1