Model
31142_ 31241_ 31162_ 31261_
Type
A1 A2 A1 A2
Cabinet Size (in feet)
4
4
6
6
Sash
Opening
(inches) 12 10 12 10
Starting Serial #
1
1303_
1303_
1303_
1303_
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
Nominal Average Inflow (FPM)
85+/-5
105+/-5
85+/-5
105+/-5
Supply HEPA Data
Labconco
P/N
3838401 3838401 3838403 3838403
Exhaust HEPA Data
Labconco
P/N
3838501 3838501 3838503 3838503
Motor/Blower Data
Labconco P/N
2
3832201
3832201
3832208
3832208
Motor HP
1/2
1/2
3/4 3/4
Fluorescent/UV Lamp Data
Fluorescent Lamps (2 each)
F32T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
UV Lamp
G30T8
G30T8
G30T8
G30T8
1.
The primary serial tag is on the lower outside edge of the right corner post.
The secondary serial tag is located on the front of the electronics module on
the top right side of the cabinet. The first two digits of the serial number are
the year of production; the next two are the month. The next 5 digits are the
sequence of production, and the letter following the serial number is the
revision level of the cabinet.
2.
Each motor must be programmed by Labconco for the appropriate width
cabinet.