Chapter 6: Maintaining The Procedure Station
Product Service 1-800-522-7658
45
Original instructions
Resetting a Circuit Breaker:
To reset any of the circuit breakers located on the left side of the electronics
module, depress the white button until it sets.
Circuit Breakers- The
front breaker protects
the electrical outlets,
the rear breaker
protects the motor and
lights.
Storage
If the Procedure Station is to be left unused for more than one month, it
should be prepared for storage.
Note: The cabinet should not be stored in areas of excess humidity or
temperature extremes. If the cabinet is moved during storage, it must be
recertified before use.
1. Close the sash completely and seal the bottom edge and the exhaust outlet
with plastic sheeting.
2. Unplug the cabinet.
3. Ensure that the cabinet will not be moved or disturbed while being stored.
Figure 6-5