This Appendix contains technical information about the Freeze Dryer including
electrical specifications and environmental operating conditions.
Electrical Specifications
• Nominal amperage – Models 7806020 & 7806021: 8A
• Nominal amperage – Models 7806030, 7806031, 7806040, 7806041,
7806060, 7806061, 7806070, & 7806071: 4A
• Frequency: All Models 50/60 Hz
• Phase: Single
Environmental Conditions
• Indoor use only.
• Maximum altitude: 6562 feet (2000 meters).
• Ambient temperature range: 41° to 104°F (5° to 40°C).
• Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
• Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
• Transient over voltages according to Installation Categories II (Over
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
• Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.