Chapter 6: Maintaining Your XT Clean Bench
Product Service 1-800-522-7658
32
Diffuser Removal:
1. Locate and remove the diffuser screws, shown in Figure 6-3. Pull the diffuser
straight out.
HEPA Filter Replacement:
The HEPA filters should only be serviced by a qualified
certification technician. Following replacement of the HEPA
filters, a qualified certification technician MUST recertify the
clean bench.
1. Unplug the clean bench.
2. Remove the diffuser as described above.
3. Loosen and remove the filter clamps and filter center strap as shown in Figure
6-4.
THE TOP, BOTTOM AND SIDE FILTER CLAMPS ARE
DIFFERENT ASSEMBLIES.
4. Gently pull out the Secondary HEPA filter gasket on the top and both sides of
the HEPA filter as shown in Figure 6-4. Save the gasket for reinstallation.
)
Figure 6-3