Chapter 6: Maintaining Your XT Clean Bench
Product Service 1-800-522-7658
30
Service Operations
Resetting the Circuit Breaker:
1. The circuit breaker is located next to the power cord on the lower right panel
of the bench as shown in Figure 6-1. If the circuit breaker trips, it can be reset
by pressing the white button in.
Changing the Prefilter:
The prefilter should be replaced at least quarterly, or more often as conditions
require. See Appendix A: XT Clean Bench Components for replacement prefilter
ordering information.
1. Make sure the clean bench is off.
2. Using a Phillips screwdriver, remove the prefilter retainer(s) located on the
front edge of the prefilters, as shown on Figure 6-1.
3. Remove both prefilters by sliding them straight forward.
4. Install new prefilters by reversing the above steps.
Do NOT contact blower wheel while still in motion.
NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.
Figure 6-1
!
Pre-Filter
Retainer
Circuit Breaker