Routine maintenance schedule – Labconco Purifie Horizontal Clean Benches 32000 Series User Manual
Page 33

Chapter 6: Maintaining Your XT Clean Bench
Product Service 1-800-522-7658
29
Routine Maintenance Schedule
Under normal operation, your Purifier XT Clean Bench will require little routine
maintenance. The following schedule is recommended:
Weekly
• Wipe down the interior surfaces of the clean bench with stainless steel
cleaner or 70% ethanol and allow to dry.
• Using a damp cloth, clean the exterior surfaces of the clean bench,
particularly the front and top of the clean bench to remove any
accumulated dust.
Monthly (or more often as required)
• Check all service valves, if so equipped, for proper operation.
• Check the prefilters and replace if necessary. The prefilters should be
replaced at least quarterly.
• All weekly activities.
Quarterly
• Replace the prefilter. See Appendix A: XT Clean Bench Components for
ordering information.
• All monthly activities.
Annually
• Have the clean bench recertified by a qualified certification technician.
• All quarterly activities.