Routine maintenance schedule, Hepa – Labconco Class I Enclosure Models 3980423 User Manual
Page 38
Chapter 6: Maintaining Your Filtered Enclosure
Product Service 1-800-522-7658
34
34
Routine Maintenance Schedule
Weekly
Wipe down the interior surfaces of the enclosure with a disinfectant or
cleaner, depending upon the usage of the unit and allow to dry.
Using a damp cloth, clean the exterior surfaces of the enclosure,
particularly the front and top to remove any accumulated dust.
Operate the exhaust system, noting the airflow velocity through the
enclosure using a source of visible smoke. Airflow monitors are
recommended for constant monitoring.
Monthly (or more often as required)
Determine the actual face velocity through the sash opening of the
enclosure where the average reading should be at the specified velocity.
(Use calibrated thermal anemometer or other approved apparatus). Airflow
alarms are recommended for constant monitoring.
The enclosure rear baffle should be checked for any blockage to ensure that
the enclosure is maintaining proper airflow.
All weekly activities.
Check face velocity. Increase speed control or change HEPA filter when
face velocity of the enclosure drops below the recommended speed for your
facility or if the airflow alarm monitor alerts you. Airflow monitors are
recommended.
While the enclosure is filled with the contaminant, test filter condition on
Odor Control carbon filters using the appropriate gas detector tube at
intervals of 20% of the total estimated time. The exception to the 20%
recommendation is formaldehyde or any carcinogen or suspected
carcinogen. These hazardous chemicals must be checked at least every
10% of the total estimated time. Gas detector tubes for the specific
chemicals that are being used in the enclosure can be obtained from your
laboratory supply dealer.
Replace Odor Control carbon filters when chemical breakthrough is
indicated by odor, time, detector tube, or for some chemicals, analytical
instrumentation. See “Replacing Odor Control Carbon Filters” section of
this manual in Chapter 6.
Annually
Replace the fluorescent lamps. Replace UV lamps if equipped.
Verify proper operation of the UV closure panel safety switch.
Have the enclosure validated by a qualified certification technician. See
Certification and Recertification in Chapter 6.
All monthly activities.
HEPA
CARBON
CARBON
- Class I Enclosure Models 3980421 Class I Enclosure Models 3980403 Class I Enclosure Models 3980401 Class I Enclosure Models 3980323 Class I Enclosure Models 3980321 Class I Enclosure Models 3980303 Class I Enclosure Models 3980301 Class I Enclosure Models 3980223 Class I Enclosure Models 3980221 Class I Enclosure Models 3980203 Class I Enclosure Models 3980201 HEPA Filtered Enclosure Models 3980422 HEPA Filtered Enclosure Models 3980420 HEPA Filtered Enclosure Models 3980402 HEPA Filtered Enclosure Models 3980400 HEPA Filtered Enclosure Models 3980322 HEPA Filtered Enclosure Models 3980320 HEPA Filtered Enclosure Models 3980302 HEPA Filtered Enclosure Models 3980300 HEPA Filtered Enclosure Models 3980222 HEPA Filtered Enclosure Models 3980220 HEPA Filtered Enclosure Models 3980202 HEPA Filtered Enclosure Models 3980200