The Purifier Enclosures were also subjected to an industrial hygiene air-monitoring
study to confirm the performance of the enclosure for powders. The Purifier
Enclosures demonstrated excellent containment when used by an operator using
excellent technique and good containment when used by an operator using
marginal technique. While no enclosure can compensate for improper technique,
these tests confirm the Purifier Enclosures provide a safe working environment.
NOTE: Face velocity profiles and smoke testing should be
performed frequently per your organization’s quality system to
ensure safe performance.
Sealing the Filtered Enclosure to the
Work Surface
When the filtered enclosure has been set in place, ducted, (if necessary), and wired,
it may be sealed at the work surface to prevent spilled materials from collecting
under the walls. A bead of silicone sealant is recommended to seal the filtered
enclosure to the work surface.
!
- Class I Enclosure Models 3980421 Class I Enclosure Models 3980403 Class I Enclosure Models 3980401 Class I Enclosure Models 3980323 Class I Enclosure Models 3980321 Class I Enclosure Models 3980303 Class I Enclosure Models 3980301 Class I Enclosure Models 3980223 Class I Enclosure Models 3980221 Class I Enclosure Models 3980203 Class I Enclosure Models 3980201 HEPA Filtered Enclosure Models 3980422 HEPA Filtered Enclosure Models 3980420 HEPA Filtered Enclosure Models 3980402 HEPA Filtered Enclosure Models 3980400 HEPA Filtered Enclosure Models 3980322 HEPA Filtered Enclosure Models 3980320 HEPA Filtered Enclosure Models 3980302 HEPA Filtered Enclosure Models 3980300 HEPA Filtered Enclosure Models 3980222 HEPA Filtered Enclosure Models 3980220 HEPA Filtered Enclosure Models 3980202 HEPA Filtered Enclosure Models 3980200