Storage – Labconco Total Exhaust Biological Safety Cabinets 30368 Series User Manual
Page 52
Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
46
Original instructions
3. Thoroughly surface decontaminate the UV lamp and the work area of the
cabinet.
4. Unplug the cabinet or turn off the System Reset Switch, located on the
top of the cabinet.
5. Remove the UV lamp by rotating it 90 degrees and lifting it straight up
and out of its sockets.
6. Install new lamp by reversing the removal procedure.
Resetting a Circuit Breaker:
To reset any of the circuit breakers located on the left side of the electronics
module, depress the white button until it sets.
Circuit Breakers- The
front breaker protects
the electrical outlets,
the rear breaker
protects the motor and
lights.
Storage
If the biosafety cabinet is to be left unused for more than one month, it should
be prepared for storage.
Note: The cabinet should not be stored in areas of excess humidity or
temperature extremes. If the cabinet is moved during storage, it must be
recertified before use.
1. Close the sash completely and seal the bottom edge and the exhaust outlet
with plastic sheeting.
2. Unplug the cabinet.
3. Ensure that the cabinet will not be moved or disturbed while being stored.
Figure 6-5