Security lock – Labconco Total Exhaust Biological Safety Cabinets 30368 Series User Manual
Page 40
Chapter 5: Using The Cabinet
Product Service 1-800-522-7658
34
Original instructions
Security Lock
Note: The Security Lock feature “locks” the keypad to prevent unauthorized
use of the cabinet. The security lock is unlocked by pressing the
▼
button
three times within 2 seconds. The lock reactivates whenever the sash is
closed, or the blower shuts off.
From the Settings menu screen, use the
▲
and
▼
buttons on the touchpad to highlight the
Security
Lock
option-it will turn white when selected. Press
OK
to enter the Security Lock screen. Using the
▲
and
▼
buttons on the touchpad, highlight either
turned on
or
turned off
-it will turn white when
selected, press
OK
. The screen will then return to
the first Settings screen.
RS-232 Output Rate
Note: This selection will only work if the optional EN/RS-232 board is
installed.
This menu option selects the rate that the RS-232 board outputs data.
Data can be output at a rate of once per second, once every 10 seconds,
once every 30 seconds, or once per minute.
From the Settings menu screen, use the
▲
and
▼
buttons on the touchpad to highlight the
RS-232
output
-it will turn white when selected. Press
OK
to enter the RS-232 output screen. Using the
▲
and
▼
buttons on the touchpad, highlight either
the data output rate preferred. It will turn white
when selected, and then press
OK
. The screen will
then return to the first Settings screen.