Labconco Total Exhaust Biological Safety Cabinets 30368 Series User Manual
Page 51
Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
45
Original instructions
1. To reinstall the panel, reverse
these steps, ensuring that the
plastic pin in the top corners of
the dress panel properly engage
the corner posts.
Changing the Fluorescent Lamps:
1. Unplug the cabinet or turn off the System Reset Switch located on the top
of the cabinet.
2. Remove the front dress panel as noted in Figure 6-3.
3. Remove the fluorescent lamps by pulling the lamp sockets straight off
each end of the lamp, and releasing both lamps from the spring clips that
secure them in place.
4. Install the new lamps by reversing the removal procedure.
Changing the Optional UV Lamp:
Note: For optimum performance, the UV lamp should be changed on an
annual basis, or as indicated by the UV lamp timer.
The UV lamp and the work area of the cabinet must be thoroughly
decontaminated before removing the lamp.
1. Start the cabinet and let it operate for 5 minutes.
2. Raise the sash to its full open position.
Figure 6-4