Labconco Total Exhaust Biological Safety Cabinets 30368 Series User Manual
Page 50
Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
44
Original instructions
black tube with an orange cap) rests behind the towel catch, allowing the
catch to contact the rear wall of the cabinet.
Sampling/Decontamination
Tube
Towel Catch Retaining
Spring
Towel Catch
Note: Upper and lower
towel catch flanges point to
the front of the cabinet, and
the bottom of the towel
catch rests against the back
wall.
Font Panel Removal and Installation:
1. Locate and remove the two Phillips
screws that secure the front panel as
shown in Figure 6-3. They are
located on the bottom corners of the
front dress panel.
2. Swing the bottom of the dress panel
out to clear the fluorescent light and
then lift the front dress panel
straight up and away from the
cabinet.
Figure 6-3
Figure 6-2