Unloading Materials and Equipment
• Objects in contact with contaminated material should be surface
decontaminated before removal from the cabinet.
• All open trays or containers should be covered before being removed
from the cabinet.
Wipe-Down
• Wipe down the interior surfaces of the cabinet with a suitable
disinfectant, or 70% ethanol, and allow to dry.
• Periodically lift the work surface and wipe down the area beneath it.
• Inspect and clean the towel catch located at the rear of the work area,
beneath the work pan.
• Dispose of rubber gloves appropriately, and have lab coat laundered
properly.
• Wash hands and arms thoroughly with germicidal soap.
Shutdown
• Lower the sash to turn off the fluorescent light and cabinet blower and
activate the UV light if appropriate.