Chapter 8: Troubleshooting
Product Service 1-800-522-7658
36
PROBLEM
CAUSE
CORRECTIVE ACTION
Lights do not work Lamp not installed properly.
Inspect lamp installation.
Lamp wiring disconnected.
Inspect lamp wiring.
Defective lamp.
Replace lamp.
Light switch is defective.
Replace light switch.
Defective electronic ballast.
Replace ballast.
Airflow monitor
malfunction
No power.
No lights.
No display.
Power supply is not plugged into proper voltage; plug in power
supply. Verify that airflow monitor interface cables are
connected. Check fuses on FilterMate or accessory light.
No audible alarm.
Alarm has been temporarily silenced using “test/reset” or “enter”
buttons.
Wrong alarm set point.
Airflow monitor was not properly adjusted. Repeat calibration
steps outlined in the User’s Manual.
Constant audible alarm.
Check airflow and calibration of airflow monitor.
Continuous alarm.
Check the face velocity of the enclosure as the airflow of the
system may have changed. If face velocity is correct, calibrate
the airflow monitor as outlined in the User’s Manual.
Monitor alarms; air way to
airflow monitor sensor is
blocked by insects, dust or
debris.
Lightly clean the airway with clean air. Be careful not to touch
sensitive electrical components.
Audible disable will not stay
ON.
An alarm condition must be continuously present before the
audible alarm can be silenced. If flow conditions fluctuate near
the alarm set point, the airflow monitor will automatically reset it.
Action should be taken to bring the enclosure airflow into proper
operating parameters or adjust the alarm set point lower.
Lift won’t operate
or malfunctions
No power.
Lift motor is not connected to a proper outlet. Connect the plug
into the four-plex electrical box located under work surface.
Intermittent operation.
Allow lift system’s thermal override to cool for 10 minutes then
adjust height.
Photoelectric eye
malfunctions.
Either adjust photoelectric eye sensitivity or replace
photoelectrical eye if it malfunctions. See replacement parts.
Hydraulic lines pinched
Replace entire lift system. See replacement parts.
- Bulk Powder Stations 3962621 Bulk Powder Stations 3962620 Bulk Powder Stations 3962602 Bulk Powder Stations 3962601 Bulk Powder Stations 3962600 Bulk Powder Stations 3962522 Bulk Powder Stations 3962521 Bulk Powder Stations 3962520 Bulk Powder Stations 3962502 Bulk Powder Stations 3962501 Bulk Powder Stations 3962500 Bulk Powder Systems 3961622 Bulk Powder Systems 3961621 Bulk Powder Systems 3961620 Bulk Powder Systems 3961602 Bulk Powder Systems 3961601 Bulk Powder Systems 3961600 Bulk Powder Systems 3961522 Bulk Powder Systems 3961521 Bulk Powder Systems 3961520 Bulk Powder Systems 3961502 Bulk Powder Systems 3961501 Bulk Powder Systems 3961500 Bulk Powder Enclosures 3963822 Bulk Powder Enclosures 3963821 Bulk Powder Enclosures 3963820 Bulk Powder Enclosures 3963802 Bulk Powder Enclosures 3963801 Bulk Powder Enclosures 3963800 Bulk Powder Enclosures 3963622 Bulk Powder Enclosures 3963621 Bulk Powder Enclosures 3963620 Bulk Powder Enclosures 3963602 Bulk Powder Enclosures 3963601 Bulk Powder Enclosures 3963600 Bulk Powder Enclosures 3963522 Bulk Powder Enclosures 3963521 Bulk Powder Enclosures 3963520 Bulk Powder Enclosures 3963502 Bulk Powder Enclosures 3963501 Bulk Powder Enclosures 3963500