Appendix c: specifications – Labconco TXE Tall XPerEnclosure 3892022 User Manual
Page 48
Product Service 1-800-522-7658
44
Appendix C:
Specifications
This Appendix contains technical information about the XPert TXE including
electrical specifications and environmental operating conditions.
Electrical Specifications
3 Amps, 115V, 60 Hz
2 Amps, 230V, 50Hz
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.