Chapter 6: Maintaining Your TXE
Product Service 1-800-522-7658
30
1. Push the ENTER button on the face of the alarm until the Set up Menu is
displayed.
2. Scroll to SET UP and hit ENTER.
3. The PASSWORD MENU displays (The Password is 0000). Press the
ENTER button repeatedly until the CAL CONFIG MENU is displayed.
4. In the CAL CONFIG MENU set for the following:
CALIBRATION CONFIG MENU
SETTINGS
DISPLAY UNITS
FPM
LOW AIR ALARM
SEE CHART ABOVE – LOW AIR ALARM
LOW AIR CUTOFF
OFF
LOW AIR CUTOFF
SEE CHART ABOVE – LOW AIR ALARM
WARNING AIR ALARM
SEE CHART ABOVE – LOW CALIBRATION SET
POINT (Set 3-5 FPM above low air alarm)
WARNING AIR RESET
3 FPM
HIGH AIR ALARM
OFF
HIGH AIR ALARM
OFF
LOWER AIR SAMPLE FLUCTUATIONS
3%
HIGHER AIR SAMPLE FLUCTUATIONS 3%
LOWER/HIGHER AIR SAMPLE
DIFFERENCE
10 FPM
WARN TO ALARM AIR TIME
10 SECONDS
ALARM TO WARN AIR TIME
3 SECONDS
SHOW AIR FLOW
ON
SHOW TIME LINE OFF = DISPLAYS
BAR GRAPH
OFF
AUDIBLE ALARM
ENABLED
SENSOR DIFFERENCE
2%
5. To complete the CAL CONFIG, be sure to enter “DONE”. You are
returned to the Main Menu.
Calibrate Alarm
1. Scroll to set up and hit ENTER, then CALIBRATE and hit ENTER.
2. The PASSWORD MENU displays (The Password is 0000). Press the
ENTER button repeatedly until the next menu is displayed.
3. Adjust the speed of the fan until the low airflow velocity is reached. Use
your thermal anemometer readings to determine the low air velocity.
4. Enter the value of your low airflow velocity using your +/- scroll keys. Be
sure the front panel is installed. Step out of the way, wait approximately 30
seconds for air currents to settle and hit ENTER. The airflow alarm will
sample the airflow.