Chapter 5: using your txe, Routine daily work procedures – Labconco TXE Tall XPerEnclosure 3892022 User Manual
Page 23
Product Service 1-800-522-7658
19
Chapter 5:
Using Your TXE
Now that the installation is completed, you are ready to use your TXE.
Read this chapter to learn about:
Routine Daily Work Procedures.
Appropriate HEPA Filter Applications, Suitability and Guidelines.
Definition of Terms.
Prohibited Acid Use.
Routine Daily Work Procedures
Planning
Thoroughly understand procedures and equipment required before
beginning work.
Start-up
Turn on exhaust system and light.
Check the baffle air slots for obstructions.
Allow the enclosure to operate unobstructed for 1 minute.
Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate. Consult your Safety Officer for
additional personal protective equipment recommendations.
Work Techniques
Doors should be closed for powder generating processes.
Avoid using techniques or procedures that disrupt the airflow patterns of
the enclosure.