Step 3 Remove the mains plug from the lamp housing power cable (if fitted) and pass the cable through
the lamp mounting pillar and cable gland.

figure 3 Lamp cable passed trough the lamp pillar


Step 4 Electrical connection of the lamp to the voltage stabilizer




115 VAC / 60 Hz Calibration Facility
The 115 VAC / 60Hz version of the CFR has all the electronic components for powering the bulb integrated
into the lamp housing. Re-fit the plug and connect directly to a suitable mains outlet socket. Connect the
voltmeter power cable to a suitable mains outlet socket.

230 VAC / 50 Hz Calibration Facility

Connect the lamp power cable to the voltage stabilization unit, as shown in figure 4. Fit the metal cover
and connect the voltage stabilizer and voltmeter power cables to suitable mains outlet sockets.












figure 4 Cable wires connected to the voltage stabilizer terminals



Warning: To prevent electric shock during installation, all electrical devices
must be disconnected from the mains power.

Label

Function

Cable

Wire color

1 in

Neutral

Mains

Blue

2 in

Phase 230 VAC

Mains

Brown

Protective Earth

Mains

Yellow/Green

Earth

Table

Yellow/Green

3 out

Neutral / Earth

Lamp

Blue and
Yellow/Green

4 out

Phase 230 VAC

Lamp

Brown

5 out

-

-

-