3.4 Calibration Form
An example of a calibration form is given below.
Test pyranometer model:
CMP 6
Serial number:
060002
Reference pyranometer model:
CMP 6
Serial number:
060001
Sensitivity:
10.00
V/Wm
-2
(SR)
Calibration performed by:
J. Smith
Date:
20 September 2006
Signature
Reference pyranometer
Test pyranometer
Position 1
Illuminated R
1980
T
2980
Darkened RE
5
TE
10
R1=R-RE
1975
T1=T-TE
2970
Position 2
Illuminated R
2100
T
2820
Darkened RE
2
TE
8
R2=R-RE
2098
T1=T-TE
2812
Test pyranometer sensitivity
)
2
1
(
)
2
1
(
R
R
T
T
SR
ST
+
+
⋅
=
14.20
V/Wm
-2
Lamp stability
0.98 <
)
2
2
(
)
1
1
(
T
R
T
R
⋅
⋅
< 1.02
0.994
OK/Rejected