Kipp&Zonen CFR Calibration Facility User Manual

Page 19

background image

CFR manual

18

3.4 Calibration Form

An example of a calibration form is given below.

Test pyranometer model:

CMP 6

Serial number:

060002


Reference pyranometer model:

CMP 6

Serial number:

060001

Sensitivity:

10.00

V/Wm

-2

(SR)


Calibration performed by:

J. Smith

Date:

20 September 2006

Signature

Reference pyranometer

Test pyranometer

Position 1

Illuminated R

1980

T

2980

Darkened RE

5

TE

10

R1=R-RE

1975

T1=T-TE

2970

Position 2

Illuminated R

2100

T

2820

Darkened RE

2

TE

8

R2=R-RE

2098

T1=T-TE

2812



Test pyranometer sensitivity

)

2

1

(

)

2

1

(

R

R

T

T

SR

ST

+

+

=

14.20

V/Wm

-2

Lamp stability

0.98 <

)

2

2

(

)

1

1

(

T

R

T

R

< 1.02

0.994

OK/Rejected