Key parts of the pgs-100 sun photometer, 1 the radiometer 1.3.2 the control unit – Kipp&Zonen PGS-100 Sun Photometer User Manual
Page 8
.
Please follow the instructions in this section carefully for the mechanical and electrical installation of the PGS-100 Sun Photometer.
Do not turn on power to the instrument until instructed to do so.
Note
Do not connect the instrument to a computer until instructed to do so.
Note
Do not turn on power to the operating computer until instructed to do so.
2.1 Included with the product
Check the contents of the shipment for completeness (see below) and note whether any damage has occurred during transport.
If there is damage, a claim should be filed with the carrier immediately. In the case of damage and/or the contents are incom-
plete, contact your local Kipp & Zonen representative or e-mail the Kipp & Zonen customer and product support department at:
[email protected]
The following items are included with the PGS-100 Sun Photometer:
The radiometer
The control unit
Connection cables
CD-ROM
Mounting rings
2.1.1 The radiometer
The radiometer contains the optical input system, solid state spectrometer with a CCD detector array, associated electronics and
temperature stabilisation components. It has a white heat shield to reduce heating by direct solar radiation, an alignment aid
for pointing the radiometer at the sun, and sockets for factory maintenance and the cable connection to the control unit. A
drying cartridge is fitted to absorb internal moisture and to prevent internal condensation that would damage the spectrometer
and optical system.
The radiometer is powered by low voltage supplied from the control unit.
.
Throughout this manual the following symbols are used to indicate to the user important information.
Danger electricity; warns of high voltage which can cause physical injury and/or damage to the equipment.
General warning about conditions, other than those caused by high voltage electricity, which may result in physical
injury and/or damage to the equipment or cause the equipment to not operate correctly.
Note
Useful information for the user
1.1 Product overview
The PGS-100 Sun Photometer uses a solid state spectrometer with a CCD detector array to measure the direct solar radiation in
the wavelength range from 350 nm (nanometers, 10
-6
m) to 1050 nm.
The instrument consists of a radiometer, a control unit and connecting cables. Operating software is supplied on CD-ROM and
must be running on a suitable computer in order for the instrument to make measurements and to store the spectral data. A
computer is not part of the PGS-100 delivery.
The radiometer must be mounted on an automatic sun tracker in order to make measurements. A mounting kit is available for
fitting the PGS-100 radiometer to a Kipp & Zonen sun tracker.
This manual provides information related to the installation, operation, maintenance, calibration, product specifications and
applications of the sun photometer.
If any questions should remain, please contact your local Kipp & Zonen representative or e-mail the Kipp & Zonen customer and
product support department at:
Please go to
www.kippzonen.com for information about other Kipp & Zonen products, or to check for any updates to this manual
or software.
1.2 Manufacturer
The PGS-100 is sold and marketed by Kipp & Zonen and is manufactured by:
Prede Co. Ltd.
Sasamoto Building,
26-8-1 Kamidaira,
Fussa-shi,
Tokyo 197-0012
Japan
www.prede.com
Instruction Manual - PGS-100 Sun Photometer
8
1.3 Key parts of the PGS-100 Sun Photometer
1.3.1 The radiometer
The drawing below shows the key parts of the radiometer:
Radiometer housing
Sun shield
Optical input window
Alignment aid
Drying cartridge
‘MAINTENANCE’ diagnostic socket (for factory service use only)
‘CONTROL A’ socket for connection to the control unit
Serial number label
1.3.2 The control unit
The drawing below shows the key parts of the control unit:
Control unit enclosure
‘CONTROL B’ socket for connection to the radiometer
‘PC’ socket for connection to the operating computer
‘POWER’ socket for connection to AC mains power
Serial number label
CONTROL A
MAINTENANCE
CONTROL B
POWER
PC
.
2.1.2 The control unit
The control unit is contained in a weatherproof enclosure. It contains power supplies, control circuits for the radiometer
temperature stabilization, interfaces to the radiometer and communication circuitry for a computer to operate the instrument.
2.1.3 Connection cables
Three cables are supplied with the PGS-100.
One is for connection between the radiometer and the control unit.
The second is for the supply of mains AC power to the control unit.
The third is for communication between the control unit and the operating computer.
.
2.1.4 CD-ROM
The CD-ROM disk contains the operating software for the PGS-100 and necessary setup files.
2.1.5 Mounting rings
The radiometer is normally supplied with two mounting rings that are used for fitting it to a Prede sun tracker. These are not
required for fitting to a Kipp & Zonen sun tracker, with the PGS-100 Mounting Kit, and can be discarded.
Note
A computer is not part of the PGS-100 delivery.
2.2 Tools required
The tools required to fit the PGS-100 radiometer to a Kipp & Zonen sun tracker are supplied with the PGS-100 Mounting Kit.
2.3 Location and support base
The location requirements are the same as for any instrument designed to be mounted on an automatic sun tracker for the
measurement of direct solar radiation. The radiometer should have an unobstructed view of the sun throughout the complete
solar arc at all times of the year.
The base support for the sun tracker must be rigid and stable and the alignment must be unaffected by weather and environmental
effects. If this is not possible, the sun tracker must be fitted with an active tracking sun sensor to compensate for movements of
the base support.
The field of view and slope angle of the PGS-100 radiometer input optics require that it must be aligned on the axis of the sun
with a pointing accuracy of 0.2 ° or better.
The ambient operating temperature range of the PGS-100 is from -10 °C to +45 °C. For operation over the range from -30 °C to
+45 °C the Low Temperature Option is required. For operation over the range from -10 °C to +60 °C the High Temperature Option
is required. Please note that both options cannot be fitted at the same time.
2.4 Mounting
It is assumed that the PGS-100 radiometer is to be fitted to a Kipp & Zonen automatic sun tracker using the optional PGS-100
Mounting Kit.
2.4.1 Mounting rings
The radiometer is normally supplied with two mounting rings (shown in 2.1.5) that are used for fitting it to a Prede sun tracker.
These are not required for fitting to a Kipp & Zonen sun tracker, with the PGS-100 Mounting Kit, and can be discarded.
2.4.2 PGS-100 mounting kit
The mounting kit comprises two brackets, clamps, and fixings; to be fitted to a side mounting plate of a Kipp & Zonen sun tracker.
Depending upon the sun tracker configuration, a side mounting plate may also be required.
2
2
1
1
3
3
4
4
4
4
4
5
5
5
2
1
3
4
5
6
6
7
7
8
8
1
2
3
4
5