beautypg.com

Specifications functional description – Hanna Instruments HI 9064 User Manual

Page 3

background image

4

5

SPECIFICATIONS

FUNCTIONAL DESCRIPTION

HI 9065
Thermo
Hygrometer

HI 9064
Hygrometer

ON/OFF

HOLD

RH

ON/OFF

°C

°F

HOLD

RH %

°C
°F

RH% 5 . . . . . . 95

5-95 %
0-60 °C

32-140 °F

1

7

8

9

3

2

5

4

2

3

6

H °C

S

H °C

S

H

S

1) Liquid Crystal Display (LCD)
2) HOLD key, to freeze reading on display
3) ON/OFF key, to turn the meter on and off
4) °F key, to display temperature in Fahrenheit (HI 9065 only)
5) °C key, to display temperature in Centigrade (HI 9065 only)
6) RH key, to display RH measurements (HI 9065 only)
7) Battery compartment
8) RH probe socket
9) Temperature probe socket with protective cap (HI 9065 only) for

HI 765 probe series (see “Accessories” section)

10) Shielded cable of RH probe
11) Low RH calibration trimmer
12) High RH calibration trimmer
13) Polypropylene probe body
14) Perforated protective cap, to prevent damages to the sensor

HI 9064

HI 9065

Range

5.0 to 95.0% RH

--

0.0 to 60.0°C / 32 to 140°F

Resolution

0.1% RH

--

0.1°C / 1°F

Accuracy

±2% RH

--

±0.4°C / ±1 °F

(for 1 year, excluding probe error)

Typical EMC

±3% RH

Deviation

--

±2°C / ±4°F

RH Calibration

Manual, 2 point

through trimmers on RH probe

Probe

HI 70608/2

HI 70605/2

Environment

0 to 50°C (32 to 122°F);

RH max 100%

Battery Type / Life

4 x 1.5V AA / approx. 500 hours of continuous use

Dimensions

196 x 80 x 60 mm

(7.7 x 3.1 x 2.4")

Weight

500 g (1.1 lb.)

This manual is related to the following products: