beautypg.com

Hanna Instruments HI 83208 User Manual

Page 39

background image

39

Manganese HR

• Add the content of one packet of HI 93709B-0 Sodium

Periodate reagent. Replace the cap and invert to mix with
gently movements for 2 minutes.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 1 minute and 30 seconds and press Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of manganese.

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of potassium permanganate (KMnO

4

) and

permanganate (MnO

4

¯).

• Press or to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Calcium above 700 mg/L
Chloride above 70000 mg/L
Iron above 5 mg/L
Magnesium above 100000 mg/L.