Iron LR

IRON LOW RANGE

SPECIFICATIONS

Range

0 to 400 μg/L

Resolution

1 μg/L

Accuracy

±10 μg/L ±8% of reading at 25 °C

Typical EMC

±1 μg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a
violet tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93746-0

TPTZ Reagent

2 packets

REAGENT SETS

HI 93746-01 Reagents for 50 tests
HI 93746-03 Reagents for 150 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Iron LR method using the procedure described in

the

Method Selection section (see page 12).

• Fill one graduated mixing cylinder up to the 25 mL mark

with deionized water.

• Add the content of one packet of HI 93736-0 TPTZ

reagent, close the cylinder and shake vigorously for
30 seconds. This is the blank.

• Fill a cuvette with 10 mL of the blank up to the mark and

replace the cap.

• Place the cuvette into the holder and close the lid.

10 mL

25 mL