Hanna Instruments HI 83209-01 User Manual
Page 27
![background image](/manuals/737588/27/background.png)
27
Chromium VI HR
CHROMIUM VI HIGH RANGE
SPECIFICATIONS
Range
0 to 1000 μg/L
Resolution
1 μg/L
Accuracy
±5 μg/L ±4% of reading at 25 °C
Typical EMC
±1 μg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
ASTM Manual of Water and Environmental Technology, D1687-92,
Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93723-0
Powder reagent
1 packet
REAGENT SETS
HI 93723-01 Reagents for 100 tests
HI 93723-03 Reagents for 300 tests
For other accessories see page 65.
MEASUREMENT PROCEDURE
• Select the
Chromium VI HR method using the procedure described in
the
Method Selection section (see page 12).
• Fill the cuvette up to the mark with 10 mL of unreacted sample and
replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of
HI 93723-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.
10 mL