beautypg.com

Hanna Instruments HI 83209-01 User Manual

Page 25

background image

25

Total Chlorine

• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette.
Powder reagents procedure
• Add 1 packet of HI 93711 DPD reagent. Replace the cap and

shake gently for 20 seconds (or 2 minutes for seawater analysis).

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the

measurement or, alternatively, wait for 2 minutes and 30 seconds
and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of total
chlorine.

Liquid reagents procedure
• To an empty cuvette add 3 drops of HI 93701A-T

DPD1 indicator, 3 drops of HI 93701B-T DPD1 buffer
and 1 drop of HI 93701C DPD3 solution. Swirl gently
to mix and immediately add 10 mL of unreacted
sample. Replace the cap and shake gently again.

• Reinsert the cuvette into the instrument.

This manual is related to the following products: