beautypg.com

Hanna Instruments HI 83211 User Manual

Page 36

background image

36

Iron HR

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively,

wait for 3 minutes and press Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in mg/L of iron.

INTERFERENCES

Interference may be caused by:
Molybdate Molybdenum above 50 ppm
Calcium above 10000 ppm (as CaCO

3

)

Magnesium above 100000 ppm (as CaCO

3

)

Chloride above 185000 ppm.