SPECIFICATIONS

Range

0.0 to 12.5 mg/L

Resolution

0.1 mg/L

Accuracy

±0.1 mg/L ±5% of reading at 25 °C

Typical EMC

±0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, DPD method. The reaction between iodine and the reagent causes a pink

tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93718-0

DPD Reagent

1 packet

REAGENT SETS

HI 93718-01 Reagents for 100 tests
HI 93718-03 Reagents for 300 tests
For other accessories see page 66.

MEASUREMENT PROCEDURE

• Select the

Iodine method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cap and add the content of one packet of

HI 93718-0 DPD reagent. Replace the cap and shake
gently for about 30 seconds to dissolve most of the reagent.

10 mL

Iodine

IODINE