Hanna Instruments HI 83211 User Manual
Page 35

35
Iron HR
SPECIFICATIONS
Range
0.00 to 5.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.04 mg/L ±2% of reading at 25 °C
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
EPA Phenantroline method 315B, for natural and treated waters.
The reaction between iron and reagents causes an orange tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93721-0
Powder Reagent 1 packet
REAGENT SETS
HI 93721-01 Reagents for 100 tests
HI 93721-03 Reagents for 300 tests
For other accessories see page 66.
MEASUREMENT PROCEDURE
• Select the
Iron HR method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet
of HI 93721-0 reagent. Replace the cap and shake
until dissolution is complete.
10 mL
IRON HIGH RANGE