Hanna Instruments HI 83212 User Manual

Page 25

background image

48

49

The analyzed data can be managed using Hanna’s product HI92000,

Windows

®

Compatible Software

.

DATA MANAGEMENT

The instrument shows clear warning messages when erroneous conditions appear and when measured values are
outside the expected range. These messages are described below.

No Light

: The light source is not functioning properly.

Light Leak

: There is an excess amount of ambient light reaching the

detector.

Inverted cuvettes

: The sample and the zero cuvettes are inverted.

Battery Low

: The battery capacity is lower than 10%.

Light Low

: The instrument cannot adjust the light level. Please check

that the sample does not contain any debris.

ERRORS AND WARNINGS

Light High

: There is too much light to perform a measurement. Please

check the preparation of the zero cuvette.

Description

Range

Method

Ammonia MR

0.00 to 10.00 mg / L

Nessler

Ammonia LR

0.00 to 3.00 mg / L

Nessler

Chlorine, Free

0.00 to 2.50 mg / L

DPD

Chlorine, Total

0.00 to 3.50 mg / L

DPD

Copper HR

0.00 to 5.00 mg/L

Bicinchoninate

Copper LR

0 to 1000

µ

g/L

Bicinchoninate

Hydrazine

0 to 400

µ

g/L

p-Dimethylaminobenzaldehyde

Molybdenum

0.0 to 40.0 mg/L

Mercaptoacetic Acid

Phosphate HR

0.0 to 30.0 mg/L

Amino Acid

Phosphate LR

0.00 to 2.50 mg/L

Ascorbic Acid

Phosphorus

0.0 to 15.0 mg / L

Amino Acid

Silica

0.00 to 2.00 mg/L

Heteropoly Blue

Silver

0.000 to 1.000 mg/L

PAN

STANDARD METHODS