Hanna Instruments HI 83212 User Manual
Page 25

48
49
The analyzed data can be managed using Hanna’s product HI92000,
Windows
®
Compatible Software
.
DATA MANAGEMENT
The instrument shows clear warning messages when erroneous conditions appear and when measured values are
outside the expected range. These messages are described below.
No Light
: The light source is not functioning properly.
Light Leak
: There is an excess amount of ambient light reaching the
detector.
Inverted cuvettes
: The sample and the zero cuvettes are inverted.
Battery Low
: The battery capacity is lower than 10%.
Light Low
: The instrument cannot adjust the light level. Please check
that the sample does not contain any debris.
ERRORS AND WARNINGS
Light High
: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.
Description
Range
Method
Ammonia MR
0.00 to 10.00 mg / L
Nessler
Ammonia LR
0.00 to 3.00 mg / L
Nessler
Chlorine, Free
0.00 to 2.50 mg / L
DPD
Chlorine, Total
0.00 to 3.50 mg / L
DPD
Copper HR
0.00 to 5.00 mg/L
Bicinchoninate
Copper LR
0 to 1000
µ
g/L
Bicinchoninate
Hydrazine
0 to 400
µ
g/L
p-Dimethylaminobenzaldehyde
Molybdenum
0.0 to 40.0 mg/L
Mercaptoacetic Acid
Phosphate HR
0.0 to 30.0 mg/L
Amino Acid
Phosphate LR
0.00 to 2.50 mg/L
Ascorbic Acid
Phosphorus
0.0 to 15.0 mg / L
Amino Acid
Silica
0.00 to 2.00 mg/L
Heteropoly Blue
Silver
0.000 to 1.000 mg/L
PAN
STANDARD METHODS