Hydrazine

HYDRAZINE

SPECIFICATIONS

Range

0 to 400 µg/L

Resolution

1 µg/L

Accuracy

±4% of full scale

Typical EMC

±2 µg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, method

D1385-88, p-Dimethylaminobenzaldehyde method. The reaction between hydrazine
and the liquid reagent causes a yellow tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93704-0

Liquid Reagent

24 drops

REAGENT SETS

HI 93704-01

Reagents for 100 tests

HI 93704-03

Reagents for 300 tests

For other accessories see page 50.

MEASUREMENT PROCEDURE

• Select the

Hydrazine method using the procedure described

in the

Method Selection section (see page 12).

• Fill one cuvette up to the mark with 10 mL of distilled water.

• Place the cap, insert the cuvette # 1 into the holder and

close the lid.

• Press the Blank function key to start adjusting the light level.

The display will show “Blank Done” when the meter is ready
to take a zero measurement.

• Fill a second cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

10 mL

# 1

10 mL

# 2

# 1

Copper LR

• Remove the cuvette.

• Add the content of one packet of HI 93747-0 Bicinchoninate.

Replace the cap and shake gently for about 15 seconds.

• Reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown prior to

the measurement or, alternatively, wait for 45 seconds and press
READ. When the timer ends the meter will perform the reading.

• The instrument displays the results in mg/L of copper.

INTERFERENCES

Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between 6
and 8.