Hanna Instruments HI 83212 User Manual

Page 17

background image

32

33

Molybdenum

MOLYBDENUM

SPECIFICATIONS

Range

0.0 to 40.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.3 mg/L ±5% of reading

Typical EMC

±0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and
the reagents causes a yellow tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93730A-0

Reagent A

1 packet

HI 93730B-0

Reagent B

1 packet

HI 93730C-0

Reagent C

1 packet

REAGENT SETS

HI 93730-01

Reagents for 100 tests

HI 93730-03

Reagents for 300 tests

For other accessories see page 50.

MEASUREMENT PROCEDURE

• Select the

Molybdenum method using the procedure

described in the

Method Selection section (see page

12).

• Fill the cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when

the meter is zeroed and ready for measurement.

• Add 12 drops of the HI 93704-0 reagent to each

cuvette. Replace the caps and shake gently to mix.

• Place the blank (#1) into the holder and close the lid.

• Press TIMER and the display will show the countdown

prior to zeroing the blank. The display will show “-0.0-”
when the meter is zeroed and ready for measurement.

• Remove the blank.
• Insert the cuvette with the reacted sample (# 2) into the

instrument and close the lid.

• Press READ to start the reading. The instrument displays

concentration in µg/L of hydrazine.

Hydrazine

x 12

# 1

# 2

x 12

# 1

# 2

INTERFERENCES

Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines