Hanna Instruments HI 95725C User Manual
Page 9
16
17
• Add the content of one packet of
HI 93722-0 reagent. Replace the cap and
shake gently for about 10 seconds
(dissolution is not complete). A milky
turbidity will form if cyanuric acid is present.
Note: accuracy is not affected by
undissolved reagent powder.
• Reinsert the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Wait for 45’, then press READ and “SIP”
will blink during measurement.
• Remove the cuvet.
• The instrument directly displays
concentration in ppm (mg/L) of cyanuric
acid on the Liquid Crystal Display.
Note: discard reacted sample immediately
after measurement, or a white rilm may
stain the glass cuvet.
INTERFERENCES
INTERFERENCES
INTERFERENCES
INTERFERENCES
INTERFERENCES
Turbidity already present in the sample may cause interference during measurement.
Cyanuric Acid
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range
0 to 80 mg/L
Resolution
1 mg/L
Precision
±2 mg/L @ 20 mg/L
Method
Adaptation of the turbidimetric method. The reaction between cyanuric acid
and the reagent causes a white suspension in the sample.
REQUIRE REAGENTS
REQUIRE REAGENTS
REQUIRE REAGENTS
REQUIRE REAGENTS
REQUIRE REAGENTS
Code
Description
Quantity/test
HI 93722-0
Cyanuric Acid Reagent
1 packet
REAGENT SETS
REAGENT SETS
REAGENT SETS
REAGENT SETS
REAGENT SETS
HI 93722-01
Reagents for 100 tests
HI 93722-03
Reagents for 300 tests
For other accessories see page 24.
CYANURIC ACID
MEASUREMENT
MEASUREMENT
MEASUREMENT
MEASUREMENT
MEASUREMENT
• Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
• Place the cuvet into the holder and ensure
that the notch on the cap is positioned
securely into the groove.
• Press ZERO/SET and “SIP” will blink on
the display.
• After a few seconds the display will show
“-0.0-”. Now the meter is zeroed and
ready for measurement.
• Select the parameter code for Cyanuric
Acid “CYS” by holding ZERO/SET (see
page 12).
Cyanuric Acid