18

19

• Press ZERO/SET and “SIP” will blink on

the display.

• After a few seconds the display will show

“-0.0-”. Now the meter is zeroed and
ready for measurement.

• Add 5 drops of HI 93710-0 reagent. Replace

the cap and mix the solution.

• Reinsert the cuvet into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Remove the cuvet.

• Press READ and “SIP” will blink during

measurement.

• The instrument directly displays the pH

value on the the Liquid Crystal Display.

pH

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS
Range

6.5 to 8.0 pH

Resolution

0.1 pH

Precision

±0.1 @ 7.2 pH

Method

Adaptation of the Phenol Red method. The reaction with the reagent causes a
yellow to red tint in the sample.

REQUIRE REAGENTS

REQUIRE REAGENTS

REQUIRE REAGENTS

REQUIRE REAGENTS

REQUIRE REAGENTS
Code

Description

Quantity/test

HI 93710-0

pH Reagent

5 drops

REAGENT SETS

REAGENT SETS

REAGENT SETS

REAGENT SETS

REAGENT SETS
HI 93710-01

Reagents for 100 tests

HI 93710-03

Reagents for 300 tests

For other accessories see page 24.

pH

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

MEASUREMENT

• Fill the cuvet with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

For most accurate results, use of a class A
laboratory pipette is recommended.

• Place the cuvet into the holder and ensure

that the notch on the cap is positioned
securely into the groove.

• Select the parameter code for pH “PH” by

holding ZERO/SET (see page 12).

pH