Specifications precision and accuracy – Hanna Instruments HI 96706C User Manual

Page 5

background image

5

SPECIFICATIONS

PRECISION AND ACCURACY

Range

0.0 to 15.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.3 mg/L ±4% of reading @ 25°C

Typical EMC Deviation

±0.2 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525 nm

Method

Amino Acid Method, adapted from Standard Method for the Examination
of Water and Wastewater. The reaction between phosphorus and the
reagents causes a blue tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode

;

after 1 hour of non-use in

calibration mode

;

with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7")

Weight

360 g (12.7 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93706A-0

Phosphorus Reagent A (Molybdate)

10 drops

HI 93706B-0

Phosphorus Reagent B (Amino Acid)

1 packet

Precision

is how closely repeated measurements

agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

is defined as the nearness of a test

result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of
6.00 mg/L phosphorus and a representative lot
of reagents, an operator obtained with a single
instrument a standard deviation of 0.2 mg/L.