Validation procedure – Hanna Instruments HI 96706C User Manual
Page 15
![background image](/manuals/737498/15/background.png)
15
Alternatively, wait for 5 minutes and just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
• At the end of measurement, the
instrument directly displays concentration
in mg/L of phosphorus on the LCD.
INTERFERENCES
Calcium (Ca
2+
):
above 10,000 mg/L as CaCO
3
Chloride (Cl¯):
above 150,000 mg/L as Cl¯
Ferrous iron (Fe
2+
):
above 100 mg/L as Fe
2+
Magnesium (Mg
2+
):
above 40,000 mg/L as CaCO
3
Sulfide (S
2
¯)
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:
Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained.
•
Turn the meter on by pressing ON/OFF.
• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
• Place the CAL CHECK
TM
Standard
HI 96706-11
Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.