Alternatively, wait for 5 minutes and just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the

instrument directly displays concentration
in mg/L of phosphorus on the LCD.

INTERFERENCES
Calcium (Ca

2+

):

above 10,000 mg/L as CaCO

3

Chloride (Cl¯):

above 150,000 mg/L as Cl¯

Ferrous iron (Fe

2+

):

above 100 mg/L as Fe

2+

Magnesium (Mg

2+

):

above 40,000 mg/L as CaCO

3

Sulfide (S

2

¯)

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

Turn the meter on by pressing ON/OFF.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

• Place the CAL CHECK

TM

Standard

HI 96706-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.