• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.

• Remove the cuvette.

• Add 10 drops of HI 93706A-0 Phosphorus

Reagent A to the cuvette.

• Add the content of one packet of

HI 93706B-0

Phosphorus Reagent B to

the cuvette. Replace the cap and shake
gently until dissolution is complete.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.