Calibration procedure – Hanna Instruments HI 96710 User Manual

Page 21

background image

21

• Press CAL CHECK and the lamp, cuvette

and detector icons together with “CAL
CHECK

” will appear on the display,

depending on the measurement phase.

• At the end of the measurement the display

will show the validation standard value.

The reading should be within specifications as reported in the CAL CHECK

TM

Standard

Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.

CALIBRATION PROCEDURE

Note:

It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or

ON/OFF

keys.

Warning

: Do not calibrate the instrument with standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

When calibrating, only the selected range is affected.

• Turn the meter on by pressing ON/OFF.

When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.

• Press and hold CAL CHECK

for three

seconds to

enter

calibration mode

. The

display will show “CAL” during
calibration procedure. The blinking
“ZERO”

asks for instrument zeroing.

• Place the CAL CHECK

TM

Standard

HI 96710-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.