• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.

Alternatively, wait for one minute and
just

press

READ/ /TIMER.

In both cases,

the lamp, cuvette and detector icons will
appear on the display, depending on the
measurement phase.

• The instrument directly displays the

concentration in mg/L of free chlorine on
the Liquid Crystal Display.

INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Chlorine Dioxide
Ozone
Oxidized manganese and Chromium
In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for

approximately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO

3

or acidity greater than 150 mg/L

CaCO

3

, the color of the sample could disappear or develop only partially. To resolve this,

neutralize the sample with diluted HCl or NaOH.