INTERFERENCES

Interference may be caused by:
Bromine

Iodine

Chlorine Dioxide
Ozone

Oxidized manganese and Chromium.

In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approxi-

mately 1 minute after adding the reagent.
In case of water with alkalinity greater than 250 mg/L CaCO

3

or acidity greater than 150 mg/L

CaCO

3

, the color of the sample could disappear or develop only partially. To resolve this,

neutralize the sample with diluted HCl or NaOH.

Alternatively, wait for 2 minutes and 30
seconds and just

press

READ/ /TIMER.

In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• The instrument directly displays the

concentration in mg/L of total chlorine on
the Liquid Crystal Display.