Specifications precision and accuracy – Hanna Instruments HI 96721C User Manual

Page 5

background image

5

SPECIFICATIONS

PRECISION AND ACCURACY

Range

0.00 to 5.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.04 mg/L ± 2% of reading @ 25°C

Typical EMC Deviation

±0.01 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525 nm

Method

Adaptation of the USEPA method 315B and Standard Method 3500-Fe B.
The reaction between iron and the phenantroline reagent causes an
orange tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode

;

after 1 hour of non-use in

calibration mode

;

with last reading reminder.

Dimensions

192 x 102 x 67 mm (7.6 x 4 x 2.6")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93721-0

Iron High Range Powder Reagent

1 packet

Precision

Precision

Precision

Precision

Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

Accuracy

Accuracy

Accuracy

Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 1.50
mg/L Iron and a representative lot of reagent, an
operator obtained with a single instrument a
standard deviation of 0.010 mg/L.