• Remove the cuvette.

• Add the content of one packet of

HI 93721-0

Iron High Range reagent to

the cuvette. Replace the cap and shake
gently until dissolution is complete.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

Alternatively, wait for 3 minutes and just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
iron on the LCD.

INTERFERENCES
Alkalinity:

above 2,000 mg/L (as CaCO

3

). May slow down

color development (negative error).
To resolve this, neutralize the sample with diluted
HCl