Validation procedure – Hanna Instruments HI 96721C User Manual

Page 15

background image

15

Calcium (Ca

2+

):

above 10,000 mg/L as CaCO

3

Copper (Cu

2+

):

does not interfere: reagent contains a masking
agent.

Magnesium (Mg

2+

):

above 100,000 mg/L as CaCO

3

Molybdate Molybdenum:

above 25 mg/L (as Mo-MoO

4

) (positive error)

Sample containing up to 50 mg/L Mo-MoO

4

may

be measured if a reading is taken after 1 minute
instead of 3 minutes.

High Iron concentrations:

up to 200 mg/L do not inhibit color development;
above 200 mg/L color development may be inhibited.

High sulfide concentrations (S

2-

):

high sulfide concentrations may interfere
To reduce sulfide: acidify the sample and boil for
about 20 minutes, neutralize and adjust volume
before measurements.

Silica (SiO

2

):

above 175 mg/L (as SiO

2

). May slow down color

development (negative error)

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

Turn the meter on by pressing ON/OFF.

• Place the CAL CHECK

TM

Standard

HI 96721-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.