Specifications precision and accuracy – Hanna Instruments HI 96750 User Manual

Page 5

background image

5

SPECIFICATIONS

PRECISION AND ACCURACY

Range

0.0 to 10.0 mg/L LR
10 to 100 mg/L MR

Resolution

0.1 mg/L LR
1 mg/L MR

Accuracy

LR: ±1.5 mg/L ±7% of reading @ 25°C
MR: ±15 mg/L ±7% of reading @ 25°C
(excluding dilution error)

Typical EMC Deviation

±0.1 mg/L LR
±1 mg/L MR

Light Source

Light Emitting Diode

Light Detector

Silicon Photocell with narrow band interference filter @ 466nm

Method

Tetraphenylborate Method. The reaction between potassium and re-
agents causes turbidity in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode;

after 1 hour of non-use in

calibration mode;

with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)

Weight

360 g (12.7 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93750A-0

Potassium Reagent

6 drops

HI 93750B-0

Powder Reagent

1 packet

Precision

Precision

Precision

Precision

Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

Accuracy

Accuracy

Accuracy

Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 2.00
mg/L potassium and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.06 mg/L.