• Fill the cuvette with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Add six drops of HI 93750A-0, replace

the cap and swirl the solution.

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will

show “-0.0-”. The meter is now zeroed
and ready for measurement.

• Remove the cuvette.
• Add the content of one packet of

HI 93750B-0

powder reagent. Replace

the cap and shake gently for about 1
minute to dissolve most of the reagent.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 2 minutes then just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
potassium on the LCD.