INTERFERENCES
The following ions do not interfere below the concentrations reported:
Ammonium

10 ppm

Calcium

10000 ppm as CaCO

3

Chlorine

12000 ppm

Magnesium

8000 ppm as CaCO

3

Sodium

8000 ppm

Color or suspended matter in large amounts will cause interference. Therefore the turbidity
should be removed by prior filtration.
For best results samples should be tested at room temperature (20

OC

C or 68

O

F). Lower

temperatures will result in higher readings.

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

Note: The validation is performed only for the selected parameter. For full validation of the
instrument, the following procedure must be performed for each parameter.

• If the analysis result is over the method range, the sample should

be diluted.

• Use the syringe to measure 10mL of sample.
• To measure exactly 10 mL of sample with the 5 mL syringe,

push the plunger completely into the syringe and insert the tip

into the sample. Pull the plunger out until the lower edge of the

seal is on the 5 mL mark of the syringe.

• Remove the cap and fill the Demineralizer Bottle with tap water.
• Replace the cap and shake gently for at least 2 minutes.
• Open the upper part of the Demineralizer Bottle cap and gently

squirt the demineralized water into the cylinder, up to the 100 mL

mark.

Note

: The ion exchange resin contained in the Demineralizer Bottle

is provided with an indicator substance. The indicator will

change from green to blue when the resin has been exhausted

and needs to be replaced.

• Pour the solution in the 170 mL beaker, replace the cap and

invert several times to mix. This is the diluted sample.

Select P2 program. The readings on diluted sample are performed

similary as for P1 method. The results are converted automatically

by meter.

Note

: The conversion factor for dilution is included in software of P2.