• Use a Checktemp or another calibrated thermometer with

a resolution of 0.1° as a reference thermometer.

• Immerse the temperature probe in

the beaker with ice and water as near
to the Checktemp as possible.

• Press and hold first CFM and then

CAL to enter the temperature cali-
bration mode.

• Execute the password procedure.

• With HI 21, the meter will ask for the calibration procedure

code number. Use or to select code 1 for the tem-
perature calibration and press CFM to enter.

• CAL will blink on the LCD until the

meter confirms a steady reading.

• When the reading has stabilized at

a point near the first calibration
point, CAL will stop blinking and an
intermittent CFM will prompt the user
to confirm the first calibration.

• If the reading stabilizes at a reading

significantly variant from the first
setpoint, an intermittent WRONG
will prompt the user to check the
beaker or baths.

• After pressing CFM the unit will switch to the second cali-

bration point.

• Select 25 or 50°C by pressing or .

0 °C

(32 °F)

°C

• When the reading has stabilized at

a point near the first calibration
point, CAL will stop blinking and an
intermittent CFM icon will prompt the
user to confirm the first calibration.

• If the display stabilizes at a value

significantly different from the first
setpoint, an intermittent WRONG
icon will prompt the user to check
and adjust the simulator and start
again.

• After pressing CFM the unit will switch to the second cali-

bration point at 350 mV.

• With HI 22 it is possible to select 1900 mV by pressing

or . After that, proceed as described above.

Note

A measure is considered stable when it varies little within a
sequence of acquisitions. The number of acquisitions is fixed
so that the waiting time for blinking “CFM” is about 20 sec-
onds.

Calibration procedure may be interrupted by
pressing CAL. If the calibration procedure is in-
terrupted this way, or if the controller is switched
off before the last step, no calibration data is
stored to EEPROM.

TEMPERATURE CALIBRATION

The pH/mV controller is factory calibrated for the mV and
temperature inputs. However, the user may also perform a
temperature calibration.
• Prepare a beaker containing ice
and water at 0°C/32°F and an-
other one with hot water at 25°C/
77°F or 50°C/122°F.

0 °C

(32 °F)

°C

50 ºC

(122 ºF)

°C